Agilent/HP 6622A 2 output 80 Watt Power Supply

Agilent/HP 6622A 2 output 80 Watt Power Supply
Manufacturer: Agilent / HP
Call for pricing

 

Agilent/HP 6622A 80 Watt Dual Power Supply

 

Agilent/HP 6622A 80 Watt Dual Power Supply

Rent me for $100 a month

   The Agilent/HP 6622A is dual-output, GPIB system power supply. The dual voltage outputs provides 20 volts at 4 amps or 50 volts at 2 amps on the high range. The linear low noise power supply outputs combine with a built in GPIB interface. Simple programming commands ease access to all features and capabilities. Dual ranging outputs provide greater flexibility in available voltage and current.    Programming is done using industry standard SCPI commands. Test system integration can be further simplified be using the VXIPlug&Play drivers. These power supplies help reduce test time with fast up and down programming, which is enhanced by an active downprogrammer which can sink the full rated current.

   Two isolated outputs are integrated into one package, conserving rack space and GPIB addresses. Dual ranges allow for more current at lower voltage levels. The outputs can be connected in parallel or series to further increase the flexibility that these products offer the system designer. Programming is done using industry standard SCPI commands and test system integration can be further simplified be using the VXIPlug&Play drivers. These power supplies help reduce test time with fast up and down programming, which is enhanced by the active down-programmer which can sink the full rated current. These power supplies are very useful on the R&D bench. The accuracy of both the programming and the measurement systems allow precise control and monitoring of prototype bias power. The extensive protection features protect valuable prototypes, including very fast CV/CC crossover. The power supply can be controlled from either the front panel keypad or, for automated testing, from the GPIB.

Features

  • GPIB programming of voltage and current
  • Measured voltage and current read-back over the GPIB
  • Two outputs can be connected in parallel or in series
  • Automatic selection of operating voltage and current range
  • Programmable over-voltage and over-current protection
  • Self-test occurs at power-on from an GPIB command
  • Electronic calibration via GPIB
  • 10 store-recall states
  • SCPI (Standard Comments for Programming Instruments)
  • VXI plug&play drivers
  • DC Floating Voltage: All outputs can be floated up to ±240 Vdc from chassis ground
  • Remote Sensing: Up to 1 V drop per load lead. The drop in the load leads is subtracted from the voltage available for the load.
  • Command Processing Time: 7 ms typical with front-panel display disabled
  • Down Programming: Current sink limits are fixed approximately 10% higher than source limits for a given operating voltage above 2.5 V
  • Input Power: 550 W max., 720 VA max.
  • GPIB Interface Capabilities: SH1, AH1, T6, L4, SR1, RL1, PP1, DC1, DT0.
  • Regulatory Compliance: Listed to UL1244; conforms to IEC 61010-1; carries the CE mark.
  • Size: 425.5 mm W x 132.6 mm H x 497.8 mm D (16.75 in x 5.22 in x 19.6 in)
  • Weight: Net, 17.4 kg (38 lb); shipping, 22.7 kg (50 lb)
  • Programming is done using industry standard SCPI commands.
  • Test system integration can be further simplified be using the VXIPlug&Play drivers.
  • These power supplies help reduce test time with fast up and down programming, which is enhanced by an active downprogrammer which can sink the full rated current.
     

US-Instrument.com