   The Agilent/HP 16048G consists of a direct attachment, 4-terminal pair type fixture which is equipped with four BNC (m) terminated-coaxial test leads. It is used on Agilent/HP 4294A LCR Meters. These test leads are used to attach user-fabricated test fixtures. DC bias levels of up to ±42V can be applied to the 16048G. Frequency response is to 110 MHz. Cable length is 1 meter.

Compensation for Fixture Residual Impedance Error

   The Agilent/HP 16048G has inherent stray capacitance, residual inductance, and residual resistance that affect the accuracy of measured values. To compensate for, or negate, these residuals to minimize measurement error, the instrument's Open/Short compensation procedure should be performed. The procedure is given in the instrument's operating manual.